| TEMPERATURE |
|
| Operating Range |
105ēC to 135ēC |
| Display Range |
5ēC to 140ēC |
| GAUGE PRESSURE |
|
| Operating Range (gauge) |
3 to 28.4 psi (0.2 to 2.0 kg/cm˛ ) |
| TIMER |
|
| Sterilization |
1 to 250 minutes |
| Drying |
1 to 90 minutes |
| DIMENSIONS |
|
| Overall (W x D x H) |
(20.1 x 24.8 x 40.9) in (51 x 63 x 104) cm |
Chamber (Diameter x Height) |
50 liter 11.8" x 28" (30 cm x 71 cm) |
| SAFETY DEVICES |
Temperature- and pressure-sensitive lid interlock, over-pressure power cutoff, over-temperature power cutoff, low-water power cutoff, mechanical pressure relief valve, lid-closure detector, over-current detector and circuit breaker with ground fault circuit interrupter.
|
| PROGRAMS |
|
| 1. For solids with drying |
Heating and air removal from chamber by vacuum pump (pre-vac) -> Sterilization -> Water and steam exhaust (fast) -> Drying by heat and vacuum -> Completion |
2. For wastes/solids without drying |
Heating -> Pre-vac -> Sterilization -> Steam exhaust (fast) -> Completion |
| 3. For liquids/agar |
Heating -> Sterilization -> Auto steam exhaust (programmable rates, or ambient cooling w/o steam exhaust) -> Completion |
4. Drying without sterilization |
Drying with heat and vacuum (time programmable) -> Completion(for additional drying after Program #1, if desired) |
| ACCESSORIES |
|
| Stainless steel |
Mesh wire baskets, buckets (pails), and test tube rack |
| POWER |
|
| Source |
120V, 50/60 Hz, single phase (220V/240V available on request) |
| Consumption |
2000 watts (2000 VA) |
| SHIPPING WEIGHT |
291 lb (133 kg) |