back to autoclaves HV-SERIES
Models HV-85 and HV-110 Specifications
 MODEL HV-85 HV-110
 TEMPERATURE  
 Operating Range 105ºC to 135ºC
 Display Range 137ºC
 Warming Range 45ºC to 60ºC
 Agar Dissolution Heating at 60ºC to 100ºC for up to 1 hour (temp. and time programmable), followed by 45ºC to 60ºC (temp. programmable) for up to 20 hours.
 GAUGE PRESSURE  
 Operating Range 3 to 30 psi (gauge), 0.021 to 0.21MPa
 Maximum
 Allowable
38 psi (gauge), 0.26MPa
 TIMER  
 Sterilization 1 to 250 minutes
 Delayed Auto-start Program autoclave to start 1 minute to 7 days later
 DIMENSIONS  
 Overall (W x D x H) (26 x 25.6 x 39.4) in
(66 x 65 x 100) cm
(26 x 25.6 x 46.5) in
(66 x 65 x 118) cm
 Chamber (Diameter x Height) 85 liters
16.5" x 24.2" (42 cm x 61.5 cm)
110 liters
16.5" x 31.3" (42 cm x 79.5 cm)
 SAFETY DEVICES Dual-sensing lid interlock, over-pressure power cutoff, over-temperature power cutoff, low-water power cutoff, mechanical pressure relief valve, microprocessor-controlled pressure relief valve, lid-closure detector, over-current detector and circuit breaker
 PROGRAMS  
 1. For agar Heating -> Sterilization -> Exhaust (slow, programmable) -> Warming -> Completion
 2. For liquids Heating -> Sterilization -> Exhaust (slow, programmable) -> Completion
 3. For solids Heating -> Sterilization -> Exhaust (fast, programmable) -> Completion
 4. For melting agar
     or for faster start-up
Rapid melting at 60°C to 100°C (temperature and time programmable) -> Maintain at 45°C to 60°C (temperature programmable) -> Completion
 ACCESSORIES  
 Stainless steel Mesh wire baskets, buckets (pails), and tube rack
 Factory or field installed Floating (load) sensor, auto water supply unit and digital printer
 POWER  
 Source 208V/220V/240V, 50/60 Hz, single phase
 Consumption 3000 watts (3000VA) 4000 watts (4000VA)
 SHIPPING WEIGHT 282 lb (128 kg) 308 lb (140 kg)
WARRANTY : Autoclaves are protected by a 12-month warranty.