| TEMPERATURE |
|
| Operating
Range |
105°C to
126°C |
| Display
Range |
5°C to 129°C |
| GAUGE PRESSURE |
|
| Operating
Range (gauge) |
3 to 22 psi
(gauge) |
| Maximum
Allowable |
38 psi (gauge),
0.26MPa |
| TIMER |
|
| Sterilization |
1 to 999 minutes,
or continuous |
| DIMENSIONS |
|
| Overall
(W x D x H) |
(17 x 21 x 42)
in (43 x 53 x 107) cm |
Chamber
(Diameter x Height) |
50 liter 11.8"
x 26.4" (30 cm x 67 cm) |
| SAFETY
DEVICES |
Over-pressure
power cutoff, over-temperature power cutoff, low-water power cutoff,
mechanical pressure relief valve, over-current detector and circuit
breaker |
| PROGRAMS |
|
| 1. For agar |
Sterilization
-> Slow Pulse Exhuast -> Warming -> Completion |
| 2. For agar |
Sterilization
-> Natural Cooling -> Warming -> Completion |
| 3. For liquids
|
Sterilization
-> Slow Pulse Exhaust -> Completion |
| 4. For Solids
|
Sterilization
-> Natural Cooling -> Completion |
| 5. For melting
agar |
Warming at
45°C to 80°C -> Completion |
| ACCESSORIES |
|
| Stainless
steel |
Mesh wire baskets,
buckets (pails), and test tube rack |
| POWER |
|
| Source |
120V, 50/60
Hz, single phase (220V available on request) |
| Consumption |
2000 watts (2000
VA) |
| SHIPPING
WEIGHT |
185 lb (84 kg) |