back to HA-300MD
HA-300MD Specifications
 TEMPERATURE  
 Operating Range 105ºC to 126ºC
 Display Range 5ºC to 129ºC
 GAUGE PRESSURE  
 Operating Range (gauge) 3 to 30 psi (gauge), 0.021 to 0.21MPa
 Maximum Allowable 38 psi (gauge), 0.26MPa
 TIMER  
 Sterilization 1 to 999 minutes, or continuous
 Drying 1 to 120 minutes
 DIMENSIONS  
 Overall (W x D x H) (17.3 x 20.5 x 42.1) in (44 x 52 x 107) cm
 Chamber
 (Diameter x Height)
50 liter 11.8” x 26.4” (30 cm x 67 cm)
 SAFETY DEVICES Over-pressure power cutoff, over-temperature power cutoff, low-water power cutoff,
mechanical pressure relief valve, over-current detector and circuit breaker
 PROGRAMS  
 1. For solids with drying Sterilization -> Manual Steam/H2O Exhaust -> Drying -> Completion
 2. For solids w/o drying Sterilization -> Auto Steam Exhaust -> Completion
 3. For liquids Sterilization -> Natural Cooling -> Completion
 4. For additional drying Drying -> Completion
 ACCESSORIES  
 Stainless steel Mesh wire baskets, buckets (pails), and test tube rack
 POWER  
 Source 120V, 50/60 Hz, single phase (220V available on request)
 Consumption 2000 watts (2000 VA)
 SHIPPING WEIGHT 180 lb (82 kg)
WARRANTY : Autoclaves are protected by a 12-month limited warranty.