HG-50 and HG-80 Specifications
| MODEL | HG-50 |
HG-80 |
| Overall (W x D x H) | 19.7" x 27.2" x 35.0" (45.5 cm x 6.91 cm x 89.0 cm) |
19.7" x 27.2" x 40.6" 45.5 cm x 69.1 cm x 103.0 cm |
| Chamber (D x H) | 50 liters 14.3" x 18.2" (36.4 cm x 47.0 cm) |
76 liters 14.3" x 28.7" (36.4 cm x 73.0 cm) |
| Temperature Range | 105°C to 135°C |
|
| Sterilization Timer | 1 to 300 minutes |
|
| Agar Warming Range | After sterilization, agar can be maintained from 45°C to 60°C (programmable) for up to 24 hours to prevent coagulation |
|
| Delayed Auto-Start | Program autoclave to start 1 to 99 hours later |
|
| Cooling Fan | Switch to turn fan on or off |
|
| Air Removal Timer | Programmable, 3 to 9 minutes after temperature has reached 97°C |
|
| Steam Venting | Selectable from: No venting and three other venting rates |
|
| Temperature for Release of Lid Interlock | Programmable, 60°C to 97°C for greater safety and flexibility |
|
| Sterilization Program Cycles | 1) Liquid cycle, 2) Agar (liquid with warming) cycle, and 3) Solid cycle |
|
| Additional Cycles for Agar Dissolution | Cycle for dissolving and keeping agar warm (no sterilization) |
|
| Agar Dissolution Protocol | Initially at 60°C-100°C for 1-60 min., then at 45°C-60°C for up to 24 hrs. |
|
| Safety Features | Circuit breaker with GFCI, low-water power cut-off, over-temp power cutoff, over-pressure power cut-off, safety valve, sensor for improper lid closing |
|
| Steam Condenser | 15-liter water tank |
|
| Drain Water Bottle | 2 liters |
|
| Power Requirements | 120V or 220V, 50/60 Hz, single phase |
220V, 50/60 Hz, single phase |
| Power Consumption | 2.0kW, optional 3.0kW for 220V only |
3.9 kW |
| Shipping Weight | 180 lb (82 kg) |
255 lb (116 kg) |